TV2 with Easy-Touch

TV2 specifications

Published figures from the product documentation. Where the catalog and the spec sheet differ, the specification is quoted here and the difference is noted.

Models
TV2-202 (four wired sensor sockets) and TV2-201 (receiver for four wireless sensors)
Display
Color Easy-Touch LCD, about 4.5 in x 2.5 in, roughly 488 x 270 pixels, 5 in diagonal
Status
QuickCheck screen: readings in green while in range, red once a limit is exceeded
Channels
Up to four sensors. A temperature and humidity sensor uses two channels, so up to eight values can be displayed
Sample rate
Wired sensors about every 4 seconds, wireless sensors about every 20 seconds
Logging interval
Selectable from once a minute to once every 24 hours
Storage
More than 80,000 samples per sensor in non-volatile memory, about 1.5 years at a 10 minute interval
Display resolution
0.1 °C or °F and 1.0 % RH on screen; finer resolution once downloaded
Temperature range
Digital sensor -40 to 125 °C (±0.3 °C). Thermistor -20 to 80 °C (±0.3 °C). Thermocouple range and accuracy depend on type
Humidity range
0 to 100 % RH (±3.0 %)
Differential pressure
±1 in wg class sensor (±0.002 in wg); the WDP2 catalog sensor is ±0.5 in wc
Sensor leads
20 ft standard, extendable with CC-25 cable to roughly 200 ft
Alarms
High and low per sensor with delay, plus power failure, low battery and missing sensor alarms
Alarm output
Internal buzzer and flashing screen; dry normally-open relay contact, fused 0.5 A at under 30 V
Remote alerts
Text and email through TView running on a PC on the same network
Connectivity
Ethernet standard; WiFi or Power over Ethernet optional; BACnet interface optional
Power
USB from a PC, or the supplied 120 to 240 VAC adaptor
Battery backup
1100 mAh lithium-ion, at least 100 hours of continued monitoring and logging
Indicators
LCD plus LEDs: green for AC power, red for battery
Security
Password-protected on-device menu; downloaded files encrypted with 128-bit encryption
Mounting
Hook-and-loop wall mount, flush mount plate, panel mount kit, or instrument case
Enclosure size
8 3/8 in x 5 in x 1 in
Weight
About 1 lb for the display; about 3 lb shipping with accessories
Operating temperature
0 to 50 °C for the display
Software
TView for Windows, included at no extra cost
Origin
Designed and manufactured in Crestwood, Kentucky

Battery backup is quoted at 100 hours throughout, matching the specification table. Older marketing pages occasionally quoted 40 or 100 hours; treat 100 hours as the figure to design around.

Need a figure that is not here?

Ask us. If it is measurable we can usually tell you, and point you at the document that says so.

Call Request a quote